INTRODUCTION
Permanent pacemaker implantation is accomplished by transvenous insertion of endocardial leads. A number of studies have demonstrated that the mechanical effect of lead fixation on the myocardium may cause local injury, as evidenced by raised levels of plasma troponin and inflammatory biomarkers1,2. It is reported that transvenous insertion of the leads is followed by a sequence of cardiac histopathological changes, leading eventually to the formation of a fibrous connective tissue scar3. In the traditional way of pacing, most patients have a very slight troponin elevation after pacemaker implantation, leading to consider that it is not harmful to cardiac function and structure4.
Physiological pacing has always been the goal that clinicians have been striving for. From the recording of His bundle electrogram5 to the success of His bundle pacing (HBP)6, physiological pacing has made significant progress. However, high pacing thresholds and low R-wave amplitudes limited the clinical application of HBP7. As the most potentially physiological pacing method, the left bundle branch area pacing (LBBAP) has been proved feasible, with stable parameters and many benefits8,9. Different from the traditional right ventricular pacing (RVP), the ventricular lead implantation in LBBAP is accomplished by the transventricular septal method. In addition, the final electrode implantation depth is 7-11 mm, which is much deeper than that of traditional pacing10. Moreover, LBBAP is not guaranteed to be in place on the first attempt, and it may take several attempts to succeed.
It is unclear what is the myocardial damage of deeper lead implantation and of more attempts in LBBAP. The severity of myocardial damage in LBBAP needs to be evaluated with clinical data. Therefore, this study aimed to observe the changes of troponin T in patients undergoing LBBAP and explore possible factors associated with greater myocardial damage in LBBAP.
METHODS
Patient selection
From March 2019 to December 2019, patients who had symptomatic bradycardia with an indication for pacemaker implantation according to 2013 ESC/EHRA Guidelines11 underwent attempts for LBBAP at Jiangsu Province People Hospital after informed written consent was obtained. Patients were excluded if they underwent cardiac resynchronization therapy or implantable cardioverter defibrillator implantation. This study was approved by the hospital ethics committee.
LBBAP procedure
LBBAP was achieved by the transventricular septal method in the basal ventricular septum and performed using the Select Secure pacing lead (Model 3830 69 cm, Medtronic Inc., Minneapolis, MN, USA), delivered through a fixed curve sheath (C315 HIS, Medtronic Inc., Minneapolis, MN, USA)12. During implantation, the tip electrode of the lead was used for unipolar pacing and recording. Briefly, the delivery sheath with the lead tip just beyond the distal part of the sheath under anteroposterior position or 30-° right anterior oblique (RAO) was first inserted into the His bundle region where the His bundle potential (HBP) was recorded. The His bundle region was referred as a marker; the initial site for LBBAP is approximately 1-1.5 cm distal to the HBP lead position in the right ventricular (RV) septum along the line between the HBP site and RV apex under 30-° RAO fluoroscopic view. Once the lead touched the right side of the septum and paced QRS morphology showed a w pattern with a notch at the nadir of the QRS in lead V1 during pacing with an output of 5 V/0.5 ms, the lead was pointed towards the left side of the septum and screwed in place. During the lead advancement procedure, paced QRS morphology and pacing impedance were closely monitored. Once paced QRS morphology showed right bundle branch block (RBBD) (usually Qr or rSR in electrocardiogram [ECG] lead V1) or near-normal QRS complex, the lead advancement was stopped. The depth of the lead inside the ventricular septum could be estimated by fluoroscopic imaging with contrast injection. Successful LBBAP was defined as unipolar paced QRS with RBBD morphology and QRS duration (QRSd) < 130 ms13. If an acceptable LBBAP could not be achieved after attempts at five locations, then the lead was placed in the left ventricular septum14.
Measurements
The levels of troponin T were recorded before operation, 12 h after operation, and 1 week after operation. Intraoperative parameters were recorded as follows: number of attempts, number of angiographies, time of 3830 lead implantation, radiation dose of 3830 lead implantation, depth of 3830 lead implantation, and initial interventricular septal thickness. Pacing threshold, R-wave amplitude, and impedance were recorded during operation and at 1 week and 1 year after operation. Echocardiogram and ECG of patients were obtained at 1 week and 1 year of the operation.
Statistics
Categorical data were presented as numbers (%), and continuous variables with normal distribution were reported as mean ± standard error (Mean ± SE). Mean supplemented by SE and median (Mean ± SE [median]) were presented for skewed distribution continuous variables. Because the assumption of normal distribution was violated for troponin T levels (KolmogorovSmirnov test), non-parametric KruskalWallis test was performed. Pearsons correlation analysis was used to examine the relationship between levels of troponin T and possibly related factors. Multiple linear regression analysis was used to explore possible factors associated with greater myocardial damage. A two-sided p < 0.05 was considered statistically significant. All analyses were performed using SPSS statistical software (version 18; SPSS, Chicago, IL).
RESULTS
Patients
In total, 138 patients who underwent attempted LBBAP for symptomatic bradycardia were enrolled in this study. Finally, LBBAP was successfully achieved in 126 patients; in the remaining 12 patients, it failed and left ventricular septum pacing was chosen. Clinical characteristics of patients with LBBAP are summarized in table 1. The indications of pacemaker implantation were sinus node dysfunction in 70 patients (56%), atrioventricular block in 48 patients (38.4%), and atrial fibrillation with long RR interval in 7 patients (5.6%). Figure 1 shows the position of 3830 leads in the interventricular septum in successful LBBAP by X-ray and echocardiogram. Figure 2 shows the developing process of pacing ECG during LBBAP and the fluoroscopic images of sheath angiography and final stage.
All | |
---|---|
N | 126 |
Age, y | 70.4 ± 1.3 |
Male, n | 68 |
CAD, n | 26 |
Hypertension, n | 72 |
DM, n | 16 |
LVEF, % | 62.0 ± 0.6 |
IVS, mm | 10.3 ± 0.2 |
Baseline QRSd, ms | 95.5 ± 2.9 |
SND, n | 70 |
AVB, n | 49 |
AF with long RR interval | 7 |
Categorical data are presented as number, and continuous data are presented as mean ± SE. CAD: coronary artery disease; DM: diabetes mellitus; LVEF: left ventricular ejection fraction; IVS: interventricular septum; QRSd: QRS duration; SND: sinus node dysfunction; AVB: atrioventricular block; AF: atrial fibrillation.
Change of troponin T
Figure 3 shows the changes of troponin T levels in patients undergoing LBBAP before operation, 12 h after operation, and 1 week after operation, with statistical differences. Compared with those before the operation, the levels of troponin T increased significantly at 12 h after operation (96.45 ± 11.07 [69.06] vs. 16.59 ± 1.84 [11.92] ng/L, p < 0.001). The levels of troponin T were significantly decreased at 1 week after operation compared with 12 h after operation (31.16 ± 4.04 [19.73] vs. 96.45 ± 11.07 [69.06] ng/L, p < 0.001), and there were no significant differences compared with levels before the operation.
Correlation and regression analyses
Pearsons correlation analysis showed that the increase of troponin T was positively correlated with the number of attempts, the number of angiographies, time of 3830 lead implantation, and radiation dose of 3830 lead implantation, but not correlated with the depth of electrode implantation and the thickness of the interventricular septum. Multiple linear regression analysis found that only the number of attempts was an independent factor related to the change of troponin T levels (Table 2).
Correlation analysis | Regression analysis | |||
---|---|---|---|---|
r value | p value | B (95% CI) | p value | |
Number of attempts | 0.548 | 0.000* | 18.329 (12.253-24.405) | 0.000* |
Number of angiographies | 0.288 | 0.002* | 6.719 (−14.380-27.817) | 0.529 |
Time of 3830 lead implantation | 0.480 | 0.000* | 0.751 (−0.274-1.777) | 0.148 |
Radiation dose of 3830 lead implantation | 0.055 | 0.802 | | |
Depth of 3830 lead implantation | 0.141 | 0.239 | | |
Thickness of interventricular septum | 0.105 | 0.312 | | |
*Statistically significant.
We explored the overall population to identify variables correlated with the increase of troponin T. Linear regression analysis was used to adjust for covariates that were found in the correlation analysis.
Safety of model 3830 lead implantation
Five patients (4.0%) had model 3830 lead perforation (electrodes penetrating through the interventricular septum into the left ventricle), and 2 patients (1.6%) had electrode dislocation (electrodes falling out of the fixed position) during the procedure. On examination, these patients did not show symptoms or signs of discomfort. At 1 week and at 1 year after the operation, no more lead perforations and dislocations occurred. Five patients (4.0%) at 1 week and 10 patients (7.9%) at 1 year after the operation had worsened tricuspid regurgitation from normal to mild, although there was no obvious restriction in leaflet motion of the lead. Threshold, R-wave amplitude, and impedance of the 3830 lead were stable and reliable during the procedure and follow-up (Table 3). Overall, the incidence of complications in this study was very low, and there were no serious complications during the 1-year follow-up.
Intraoperative | 1 week post-operative | 1 year post-operative | |
---|---|---|---|
Lead perforation (n) | 5 | 0 | 0 |
Lead dislocation (n) | 2 | 0 | 0 |
Worsened tricuspid regurgitation (n) | - | 5 | 10 |
Interventricular septal motion abnormality(n) | 0 | 0 | 0 |
Threshold (v at 0.5ms) | 0.56 ± 0.11 | 0.54 ± 0.12 | 0.75 ± 0.25 |
R-wave amplitude (mv) | 12.5 ± 2.6 | 13.7 ± 3.5 | 10.3 ± 3.7 |
Impedance (Ω) | 772 ± 15 | 656 ± 22 | 706 ± 36 |
Comparison with other types of ventricular pacing
By retrieving related reports, we compared the troponin T change between RVP in a previous study4 and LBBAP in this study. As shown in table 4, the myocardial damage of LBBAP was more serious than that of RVP, and the ratio of troponin T levels exceeding normal range (> 30 ng/L), or the range of minimal myocardial damage (> 90 ng/L) was both higher than that of RVP. No literature was retrieved on myocardial damage in HBP.
LBBAP | RVP | |
---|---|---|
Troponin T (ng/L) | Baseline: 11.9 (7.3-20.0) | Baseline: 12 (8-17) |
Post-operative: 69.1 (37.3-143.5) | Post-operative: 32 (22-50) | |
Post-operative > 30 ng/L | 84.3% | 55.6% |
Post-operative > 90 ng/L | 37.3% | 7.8% |
Continuous data are presented as median (25-75th percentiles). LBBAP: left bundle branch area pacing; RVP: right ventricular pacing.
Figure 4 shows that in this study, a patient with a dual chamber pacemaker who previously underwent right ventricular apical pacing (RVAP) was treated by pacemaker replacement using LBBAP due to battery depletion. The original ventricular electrode was abandoned in the right ventricle without being pulled out. As shown in the figure, LBBAP, HBP, and RVAP were quite different in pacing ECG and fluoroscopic images. In the pacing ECG, LBBAP shows as an incomplete RBBD with narrow QRS, HBP as an approximately normal ECG, while RVAP as a complete left bundle branch block with wide QRS.
DISCUSSION
This study aimed to observe myocardial damage in LBBAP, a new physiological pacing way, through analyzing the changes of troponin T levels in patients undergoing LBBAP. We found that the level of troponin T increased significantly at 12 h after operation and returned to pre-operative levels in 1 week. By retrieving and comparing related reports, we found that the myocardial damage of LBBAP was more serious than that of RVP.
The results above suggest that the myocardial damage of this new pacing mode is clearly existing. The previous studies have shown that the damage caused by traditional RVP is not clinically significant15. Different from the traditional RVP, the ventricular lead of LBBAP needs to penetrate into the interventricular septum to reach the left ventricular endocardium16. The ventricular lead of LBBAP into the myocardium is much deeper than that of RVP, and it needs more attempts to succeed, which we think is the main cause of myocardial damage. However, at 1 week after operation, troponin T returned to the pre-operative level, and no obvious interventricular septal motion abnormality was found in the follow-up echocardiography. These results indicated that the myocardial damage was transient and recoverable, without damage to myocardial contraction.
As the myocardial damage was clear, it was important to find the related factors to reduce the damage. Then, correlation and regression analyses showed that only the number of attempts was an independent factor related to the change of troponin T. Therefore, achieving success with the least number of attempts is an effective way to reduce myocardial damage. The previous studies have pointed out that the ventricular nine partition method17 and typical bundle branch block morphology18 are effective to predict the success of LBBAP. Besides, intracardiac echocardiogram was an efficient tool to guide LBBAP19. The above methods or indicators increase the probability of success and decrease the number of attempts, thereby reducing myocardial damage. Of course, in some cases which are difficult to correct or succeed, other alternatives such as left ventricular septum pacing should be chosen to avoid serious myocardial damage or complications caused by too many attempts20.
In terms of complications, there were no lead perforation, lead dislocation, or interventricular septal motion abnormality during a 1-year follow-up. Mild tricuspid regurgitation occurred in 10/126 patients, not more than that seen in conventional RVP21. Besides, threshold, R-wave amplitude, and impedance were stable and reliable during the procedure and follow-up.
There are some limitations in this study. First, this is a single-center study with a small sample size and without control groups. Besides, the above results did not eliminate the influence of atrial electrode implantation. Finally, longer follow-up is needed to consolidate the conclusions of this study.
In conclusion, acute myocardial damage of LBBAP was clinically significant, which seemed more serious than RVP. There were few complications during 1 year of follow-up. The number of attempts was an independent factor related to the myocardial damage.