Intrumentación
Empleo de detector semiconductor de cSi del tipo microbandas en la obtención de imágenes radiográficas digitales de maniquíes y muestras biológicas de mamas
A. Leyvaa*, A. Cabalª, I. Piñeraª, Y. Abreuª, C.M. Cruzª, L.M. Montañob, C.C. Díazb, M. Fontaineb, C.M. Ortizc, F. Padillad y R. de la Morae
ª Centro de Estudios Avanzados y Desarrollo Nuclear, C. Habana, Cuba, email: aleyva@ceaden.edu.cu
b Centro de Investigaciónes y Estudios Avanzados del IPN, D.F., México.
c Clínica de Mama y Ginecooncología,Hospital General Tacuba, ISSSTE, D.F., México.
d Instituto Superior de Tecnologías y Ciencias Aplicadas, C. Habana, Cuba.
e Centro de Control Estatal de Equipos Médicos, C. Habana, Cuba.
Recibido el 9 de octubre de 2009
Aceptado el 9 de junio de 2009
Resumen
El presente trabajo sintetiza los resultados experimentales obtenidos en la caracterización de un detector de silicio cristalino de 64 microbandas diseñado para experimentos de física de las altas energías, con el objetivo de estudiar su posible aplicación en la imagenelogía médica, específicamente en mamografía y angiografía digital. La Investigación incluyó la obtención de radiografías bidimensionales de un maniquí mamográfico empleando el método de barrido y su comparación con similares resultados simulados matemáticamente para diferentes fuentes de rayos X. También se presentan las imágenes reales obtenidas luego de radiografiar dos muestras de tejido biológico tomadas de biopsias de mamas, donde se identifica la existencia de un tipo de lesion pretumoral. Los resultados alcanzados apuntan positivamente hacia la posibilidad real de que estos detectores de avanzada puedan ser satisfactoriamente introducidos en la obtención de imágenes radiográficas digitales médicas.
Descriptores: Detector semiconductor; radiografía digital; imagen médica; rayos X; caracterización; maniquí; biopsia.
Abstract
The present work synthesizes the experimental results obtained in the characterization of 64 microstrips crystalline silicon detector designed for experiments in high energies physics, with the objective of studying its possible application in advanced médical radiography, specifically in digital mammography and angiography. The research includes the acquisition of twodimensional radiography of a mammography phantom using the scanning method, and its comparison with similar images simulated mathematically for different X rays sources. The paper also shows the experimental radiography of two biological samples taken from biopsies of mammas, where it is possible to identify the presence of possible pathological lesions. The results reached in this work point positively toward the effective possibility of satisfactorily introducing those advanced detectors in médical digital imaging applications.
Keywords: Semiconductor detector; digital radiography; médical imaging; Xray; characterization; phantom; biopsy.
PACS: 87.59.Hp; 85.30.De; 29.40.Wk
DESCARGAR ARTÍCULO EN FORMATO PDF
Agradecimientos
Los autores quieren expresar el agradecimiento por el soporte financiero brindado por los programas Alfa de la CE, CONACYT de México y PRN de la AENTA de Cuba.
Referencias
1. J.P. Moy, Nucl. Instr. and Meth. A 442 (2001) 26. [ Links ]
2. S. Midgley et al., Nucl. Instr. and Meth. A 573 (2007) 129. [ Links ]
3. E.H.M. Heijne, Nucl. Instr. and Meth., A 571 (2007) 7. [ Links ]
4. U.L. Olsena et al., Nucl. Instr. and Meth. A 576 (2007) 52. [ Links ]
5. J.F. Condelesa, T. Ghilardi Nettob y M. Mulato, Nucl. Instr. and Meth. A 577 (2007) 724. [ Links ]
6. D. Bollini et al., Nucl. Instr. and Meth. A 515 (2003) 458. [ Links ]
7. Sistema de códigos "Origin v. 7.0", OriginLab Corporation, www.OriginLab.com [ Links ]
8. J.S. Hendricks et al.,MCNPXTM (version 2.6.B, Los Alamos National Laboratory Report, LAUR063248, 2006). [ Links ]
9. Catalogue of Diagnostic XRay Spectra & other Data (Report 78, IPEM, York, 1997). [ Links ]
10. C. Avila et al., Médical Physics 32 (2005) 3755. [ Links ]