Introduction
About 90% of blind people in the world live in countries with emerging economies and more than 80% are over 50 years of age. According to an epidemiological study on cataract carried out in our country, it was identified that in the northern part of the country, the prevalence of cataract in people over 50 years of age is 1.53% and 2.3% in the southern part of the country1. Cataract surgery is the most frequent ophthalmological surgery, and to achieve the desired refractive result, biometric measurements of the eye are crucial for the proper calculation of the intraocular lens (IOL)2. The challenge of properly calculating IOL power after refractive surgery such as radial keratotomy (RK) remains unresolved after two decades3. At present, the number of eyes that have had previous refractive surgery before cataract surgery has increased. Perhaps, even more important than proper biometry of the human eye is employing the appropriate calculation scheme to improve refractive outcomes and predictability after cataract surgery4. Instruments for measuring the eyeball have been constantly evolving, initially using ultrasound by contact or immersion technique5. Subsequently, non-contact optical biometry was used and became very popular as an easy, fast, and accurate method of assessing axial length6. The IOLMaster 500 (Carl Zeiss Meditec, Germany) is a non-contact optical biometer based on dual-beam partial coherence interferometry, which works by reflecting an infrared laser with a wavelength of 780 nm from the cornea to the retinal pigment epithelium6. Biometry based on partial coherence interferometry has been positioned as the gold standard for its high accuracy and repeatability. This technology is considered less operator dependent compared to biometers based on contact flattening7.
Automated keratometry is used to measure anterior corneal curvature. Six points of light are projected onto the cornea in a hexagonal pattern within a 2.3 mm diameter optical zone. A refractive index of 1.3375 is used to convert the radius of anterior curvature (mm) into a corneal power in diopters (D)8. As biometers have evolved, so have calculation schemes too. Formulas are classified based on the method used to determine IOL power and by the number of biometric variables used for calculation9. According to the method used, they are divided into two groups: using the principle of paraxial approximation or Gaussian optics and using the pathway of light in the optical system of the eye, also called geometric optics or ray optics10. Vergence-based IOL calculation formulas have traditionally been classified by generation from the first generation up to the fourth generation. There are formulas that work differently and are outside this generational pattern, such as ray optics and artificial intelligence formulas11. The use of formulas such as Barrett II, Haigis, Hoffer H-5, Holladay 2, and Olsen has brought a new level of accuracy and has allowed cataract surgery to become a refractive procedure10. The Panacea formula developed by David Flikier is a vergence formula that has the characteristic of including the value of the anterior and posterior corneal curvature12, thus determining the actual measurement of the anterior and posterior corneal surface to obtain the total corneal astigmatism13. This formula can be accessed from the Panacea IOL and Toric calculator software (www.panaceaiolandtoriccalculator.com)13. The Barrett Universal II formula estimates posterior corneal surface power based on anterior corneal surface measurements13. The Barrett True-K formula (true keratometry) is based on the Barrett Universal II formula, which calculates a keratometry (K) that is modified for patients with previous LASIK refractive surgery, PRK, or RK. It requires keratometric measurements before and after laser refractive treatment. In addition, the formula can predict lens power when no Barrett True-K No History formula is available for eyes with no historical data14. This formula can be accessed from the Asia-Pacific Association of Cataract and Refractive Surgeons (www.apacrs.org) and the American Society of Cataract and Refractive Surgery (www.ascrs.org) websites15. Even with the development of biometers and highly accurate formulas, calculating the power of a lens in patients undergoing RK is a challenge, as corneas are often irregular with multiple curved and flat areas, making it difficult to determine the meridian axis and refractive power16. RK alters the anterior corneal curvature and posterior anterior relationship in a different way than that caused by laser refractive surgery. The optical zone is usually < 3 mm, which means that standard keratometry taken by the biometer will measure the region representing the inflection point between the incised cornea and the indirectly flattened central zone. This leads to an overestimation of corneal power and consequently to a risk of hypermetropic refractive error after cataract surgery17. Reduction of astigmatism during cataract surgery can result in a significant improvement in visual quality. However, induced or residual astigmatism after surgery decreases visual acuity and visual quality. Therefore, reliable correction of corneal astigmatism is of utmost importance18. Cataract is the most common cause of reversible blindness in the world. The General Hospital of Mexico is a hospital of national reach with institutional programs that favor access to cataract surgery for elderly people from different parts of the country. At present, patients have high expectations of visual recovery after surgery, as they look for a significant visual quality that allows them to regain independence and avoid the use of glasses for most of their activities, which has become feasible thanks to the use of biometers and formulas that allow a very accurate calculation. Patients with a history of RK usually have poor results due to corneal instability and difficulty in obtaining the effective lens position. The aim of the study was to determine the difference in refractive error after surgery between the Barrett True-K No History and Panacea formulas, for the calculation of IOL in cataract patients with a history of RK, at the General Hospital of Mexico from March 1, 2017, to August 30, 2020.
Materials and methods
An observational, analytical, cross-sectional, and retrospective study was conducted. We reviewed the records of patients who underwent cataract surgery by phacoemulsification plus IOL fitting, model and brand of lens provided at the hospital: AcrySof IQ SN60WF, AcrySof SN60AT, and AcrySof MA60AC by Alcon, Tecnis ZCB00 and Tecnis Z9000 by Johnson and Johnson. Calculation was performed using the Barrett True-K No History formula and the Panacea formula in the anterior segment microsurgery clinic of the ophthalmology department. GPower statistical software version 3.1.9.2 was used, taking into account, the sample size calculation based on differences between means of two independent groups and a two-tailed alpha zeta, effect size of 0.8, 95% power, and significance level of 5%. A sample size of 23 files was obtained for each formula. Records of patients between 45 and 80 years of age were included, who had been operated by phacoemulsification technique with post-operative refraction record, and underwent IOL calculation using IOLMaster 500 and Barrett True-K No History and Panacea formulas, without associated ophthalmologic pathology. Patients with infectious or inflammatory keratopathy, corneal opacity, retinal pathology, patients who underwent penetrating keratoplasty, records of patients with cataract extraction requiring another surgical technique, and use of other biometry methods for IOL calculation were excluded from the study. The variables were analyzed using the IBM SPSS V.22 statistical package for Windows. The result of the predicted spherical equivalent was subtracted from the actual spherical equivalent after surgery to obtain the refractive error for each formula, and measures of central tendency were performed. Student's t-test for independent groups was used to detect the difference in means, with an effect size of 0.8.
Results
A total of 22 eyes, from 15 patients, eight males and seven females, with a mean age of 57 ± 7 years (range 47-74 years) were analyzed. Mean axial length was 25.24 ± 1.18 mm (range 22.87-26.93 mm), mean keratometry was 40.88 ± 4.55 diopter (range 32.79-51.20 D), mean anterior chamber depth was 3.90 ± 0.74 mm (range 3.21-5.20 mm), mean white-to-white distance was 10.65 ± 0.69 mm (range 10.1-12.5 mm), and mean lens thickness was 4.34 ± 0.68 mm (range 3.51-6.68 mm) (Table 1).
Parameters | Mean + SD | Range |
---|---|---|
Age | 57 ± 7 years | 47-74 years |
AL | 25.24 ± 1.18 years mm | 22.87-26.93 mm |
MK | 40.88 ± 4.55 D | 37.29 a 51.20 D |
ACD | 3.90 ± 0.74 mm | 3.21-5.20 mm |
LT | 4.34 ± 0.68 mm | 3.51-6.68 mm |
C | 10.65 ± 0.69 mm | 10.1-12.5 mm |
AL: axial length; MK: mean keratometry; ACD: anterior chamber depth; LT: lens thickness; WTW: white-to-white distance.
The lenses implanted were AcrySof IQ SN60WF in five eyes, AcrySof SN60AT in seven eyes, and AcrySof MA60AC by Alcon in one eye, Tecnis ZCB00 in eight eyes and Tecnis Z9000 by Johnson and Johnson in one eye. A mean IOL power of + 20.74 ± 5.32 (range + 12.22-+ 29.50 D) was implanted with Barrett True-K, while with Panacea, the means IOL power implanted was + 20.10 ± 4.65 (range + 12.00-+ 28.50 D) (Table 2).
Formulas | IOL Power + SD | Range of LIO + SD powers | Refractive error + SD | Predicted spherical equivalent + SD |
---|---|---|---|---|
Barrett True-K No History | 20.74 ± 5.32 D | 12.22 + 29.50 D | 0.16 ± 1.12 D | 0.92 ± 0.81 D |
Panacea | 20.10 ± 4.65 D | 12 + 28.50 D | 0.93 ± 1.22 D | 0.18 ± 1.38 D |
The total number of eyes with the Barrett True-K No History formula that had a predicted spherical equivalent of < 1 D was 13 (59.09%). Regarding the Panacea formula, the total number of eyes that had a spherical equivalent of < 1 D was 14 eyes (63.63%), with a spherical equivalent of < 3 D was 100% of eyes with the Barrett True-K No History formula and 95.45% of eyes with Panacea formula (Fig. 1).
The total number of eyes with a refractive error of < 1 D was 15 (68.18%) with the Barrett True-K No History formula and 13 (59.09%) with Panacea formula, with a 3 D error, 100% of the eyes with the Barrett True-K No History formula and 95.45% of the eyes with Panacea formula (Fig. 2).
The predicted spherical equivalent with the Barrett True-K No History formula was −0.92 ± 0.81 and the predicted spherical equivalent with Panacea formula was −0.18 ± 1.38, p = 0.708. Regarding the formula's accuracy, the error with the Barrett True-K No History formula was −0.16 + 1.12, and the error with the Panacea formula was −0.93 + 1.22. p = 0.346 (Table 2).
Discussion
At present, determining IOL power in patients with a history of RK remains a challenge.
This has led to several proposals to solve the problem, Lyle and Jin19 advised subtracting 1.00 D from the keratometry. Chen et al. advised a refractive target of −1.50 D for post-operative refraction20.
New technologies make it possible to estimate lens power more accurately. In 2003, Aramberri et al.21 introduced the double-K method to improve accuracy for calculating the effective lens position in patients with a history of previous corneal refractive surgery. In 2007, Awwad et al.22 published findings of the double-K adjusted Holladay 1 formula in patients undergoing RK surgery, achieving good prediction of IOL power. In 2016, Abulafia et al.14 compared the accuracy of the Barrett True-K formula of the ASCRS calculator in patients with a history of myopic LASIK/PRK refractive surgery with other available methods and found that the Barrett's calculation was equal to or better than alternative methods for IOL selection.
Turnbull et al.17 compared seven formulas in 2020 for IOL calculation in cataract surgery patients with a history of RK, with the best results in those using the Barrett True-K formula.
In 2020, Dawson et al. published a comparison of refractive error between the Barrett True-K No History formula and intraoperative aberrometry in eyes with a history of RK, reporting that the total number of eyes with a refractive error of + 1 D was 38 (80.8%) of the sample for both methods23.
The aim of this study was to compare refractive error between two formulas, the well-known Barrett True-K No History and the novel Panacea, in cataract patients with a history of RK. The total number of eyes with a refractive error of < 1 D was 15 (68.18%) with the Barrett True-K No History formula and 13 (59.09%) with Panacea formula. Neither of the formulas achieved a refractive error of < 1 D after surgery in 70% of the cases. When Student's t-test was performed, no difference was found in the predicted spherical equivalent or in the refractive error when comparing both formulas regarding the total number of eyes.
This may be because the Panacea formula does not mathematically calculate the posterior corneal surface from the anterior surface, but takes the actual values of the anterior and posterior surface to provide the total corneal astigmatism, which generates more accurate calculations13.
The Barrett True-K No History formula requires, in addition to biometric data, the patient's current refraction to be input. Another condition to use it is to have a corneal topography with Atlas or Pentacam equipment or RTVue or Avanti XR optical coherence tomography, to be able to calculate the lens power. However, the Panacea calculator does not require these topographers or the use of optical coherence tomography to be able to calculate the lens power.
Regarding the biometric data, when entering the values in the respective form, we found that there are certain limitations in the range of values accepted by both calculators. The Barrett True-K formula has a limitation in the size of the anterior chamber depth, not accepting values that exceed 5 mm. Conversely, the Panacea formula does allow so because it is a formula that uses three standard deviations of the mean, which allows a wide range of values to be entered24]. With respect to lens thickness, the situation is reversed. The Panacea formula does not allow the inclusion of values exceeding 6 mm in length while Barrett True-K No History does.
Conclusion
There is no statistically significant difference comparing these two formulas, which demonstrates that Panacea is as effective as Barrett True-K No History in providing error and predicted spherical equivalent results in cataract patients with a history of RK. Although no difference in refractive error and predicted spherical equivalent outcomes was found between the Barrett True K No History and Panacea formulas in this study, further studies with a langer sample size are needed to corroborate these results and evaluate the differences between theses two formulas. Further studies with a larger sample size are needed to corroborate these results and evaluate the differences between these two formulas.