Services on Demand
Journal
Article
Indicators
Related links
Similars in SciELO
Share
Revista de investigación clínica
On-line version ISSN 2564-8896Print version ISSN 0034-8376
Abstract
TORRE-BOUSCOULET, Luis; CHAVEZ-PLASCENCIA, Elizabeth; VAZQUEZ-GARCIA, Juan Carlos and PEREZ-PADILLA, Rogelio. Precision and accuracy of "a pocket"pulse oximeter in Mexico City. Rev. invest. clín. [online]. 2006, vol.58, n.1, pp.28-33. ISSN 2564-8896.
Background. Pulse oximeters are frequently used in the clinical practice and we must known their precision and accuracy. The objective was to evaluate the precision and accuracy of a "pocket" pulse oximeter at an altitude of 2,240 m above sea level. Methods. We tested miniature pulse oximeters (Onyx 9,500, Nonin Finger Pulse Oximeter) in 96 patients sent to the pulmonary laboratory for an arterial blood sample. Patients were tested with 5 pulse oximeters placed in each of the fingers of the hand oposite to that used for the arterial puncture. The gold standard was the oxygen saturation of the arterial blood sample. Results. Blood samples had SaO2 of 87.2 ± 11.0 (between 42.2 and 97.9%). Pulse oximeters had a mean error of 0.28 ± 3.1%. SaO2 - (1.204 x SpO2) - 17.45966 (r = 0.92, p < 0.0001). Intraclass correlation coefficient between each of five pulse oximeters against the arterial blood standard ranged between 0.87 and 0.99. HbCO (2.4 ± 0.6) did not affect the accuracy. Conclusions. The miniature oximeter Nonin is precise and accurate at 2,240 m of altitude. The observed levels of HbCO did not affect the performance of the equipment. The oximeter good performance, small size and low cost enhances its clinical usefulness.
Keywords : Oximeter; Accuracy; Altitude; Calibration.