SciELO - Scientific Electronic Library Online

 
vol.58 issue3Neutron absorbed dose in a pacemaker CMOSRelative intensities of gadolinium L X-Rays, induced by proton bombardment at energies between 200 keV to 750 keV author indexsubject indexsearch form
Home Pagealphabetic serial listing  

Services on Demand

Journal

Article

Indicators

Related links

  • Have no similar articlesSimilars in SciELO

Share


Revista mexicana de física

Print version ISSN 0035-001X

Abstract

COELLO, E.A. et al. Position sensitive detection system for charged particles. Rev. mex. fis. [online]. 2012, vol.58, n.3, pp.198-204. ISSN 0035-001X.

The position sensitive detection system presented in this work employs the Anger logic algorithm to determine the position of the light spark produced by the passage of charged particles on a 170 x 170 x 10 mm3 scintillator material (PILOT-U). The detection system consists of a matrix of nine photomultipliers, covering a fraction of the back area of the scintillators. Tests made with a non-collimated alpha particle source together with a Monte Carlo simulation that reproduces the data, suggest an intrinsic position resolution of up to 6 mm is achieved.

Keywords : Scintillating charge particle detector; position sensitive; Anger logic.

        · abstract in Spanish     · text in English

 

Creative Commons License All the contents of this journal, except where otherwise noted, is licensed under a Creative Commons Attribution License