Services on Demand
Journal
Article
Indicators
- Cited by SciELO
- Access statistics
Related links
- Similars in SciELO
Share
Revista mexicana de física
Print version ISSN 0035-001X
Abstract
YAVUZKANAT, N.; ŞENYIğIT, M. and KAPLAN, M.. Investigation of the gamma-ray efficiency for various scintillation detector systems. Rev. mex. fis. [online]. 2022, vol.68, n.4, 041201. Epub May 19, 2023. ISSN 0035-001X. https://doi.org/10.31349/revmexfis.68.041201.
The calibrations of the full-energy peak efficiency, energy, and energy resolution are the main factors for successful quantitative gamma-ray analysis. The first purpose of this work is to present the calibration factors for 2”x2” NaI(Tl). The detector was irradiated with radioactive sources (137Cs, 60Co, 152Eu point sources and 211Am, 22Na cylinder-shaped sources) that emit photons at different energies from 59 keV to 1408 keV. The detector efficiency was also evaluated using Geant4 based GATE simulation. The analytical equations for efficiency calibration were found out with their parameters. Since the results obtained from the experimental and simulation were compatible with each other and with literature results, then the simulation model was modified into the novel scintillation detector systems such as LaBr3, GAGG(Ce) and SrI2.
Keywords : Efficiency; NaI(Tl); LaBr3; GAGG(Ce); SrI2; GATE.