SciELO - Scientific Electronic Library Online

 
vol.85 issue10Pregnancy rates in the first cycle of intrauterine insemination with gonadotropin stimulation in relation to number of follicles and age author indexsubject indexsearch form
Home Pagealphabetic serial listing  

Services on Demand

Journal

Article

Indicators

Related links

  • Have no similar articlesSimilars in SciELO

Share


Ginecología y obstetricia de México

Print version ISSN 0300-9041

Abstract

OVIEDO-CRUZ, H; REYES-MENDOZA, MA  and  CORTES-MARTINEZ, MA. Auscultation and automated electronic blood pressure device interchangeability in non-hypertensive pregnant women. A pilot study. Ginecol. obstet. Méx. [online]. 2017, vol.85, n.10, pp.653-658. ISSN 0300-9041.  https://doi.org/10.24245/gom.v85i10.1427.

OBJECTIVE:

To define the sample size needed to evaluate exchangeability between the auscultation method and the automated electronic device for MAP measurement in pregnancies without hypertension.

MATERIALS AND METHODS:

A standardized blood pressure measurement on both arms, and by both methods was performed on 30 pregnancies without hypertension; the combined MAP correlation coefficient was calculated, using the lower CI95% for final sample size needed.

RESULTS:

Correlation between methods was r = 0.557 (CI95% = 0.246; 0.763, p = 0.001 and r2 = 0.3097).

CONCLUSIONS:

The statistically significant correlation between methods could be clinically unacceptable for exchangeability; a sample size of n = 127 is needed to prove it.

Keywords : (MeSH): Blood Pressure Determination; Sphygmomanometers; Blood Pressure Monitors; Reproducibility of Results; Humans, Pregnancy; Preeclampsia.

        · abstract in Spanish     · text in Spanish     · Spanish ( pdf )