SciELO - Scientific Electronic Library Online

 
vol.29 issue4Environmental assessment of water-based paints for building outdoor modified with graphene oxideEffect of an electric toothbrush and bleaching tooth pastes on dental enamel. In vitro study author indexsubject indexsearch form
Home Pagealphabetic serial listing  

Services on Demand

Journal

Article

Indicators

Related links

  • Have no similar articlesSimilars in SciELO

Share


Superficies y vacío

Print version ISSN 1665-3521

Abstract

GUTIERREZ FUENTES, R.; CAMACHO LOPEZ, M.A.; JIMENEZ PEREZ, J.L.  and  CORREA PACHECO, Z.N.. Implementation of the Z-scan technique using NI cRIO 9074 system. Superf. vacío [online]. 2016, vol.29, n.4, pp.112-115.  Epub Nov 06, 2020. ISSN 1665-3521.

This paper presents the implementation of Z-scan technique using NI cRIO 9074 system to characterize different types of nanofluids and films. This technique was implemented in the Universidad Autonoma del Estado de Mexico. For this experiment a chassis NI cRIO9074 of National Instruments, a linear translation stage NRT150E and a stepper motor controller BSC203 were used, both of Thorlabs. Three steps were followed for this implementation first, the connection between NI cRIO9074 and BSC203 controller. Second, the program on LabVIEW was developed and finally, all optic part of the z-scan technique was implemented. This implementation can be used to characterize relatively thin (< 5mm) optical materials. The system testing was done with gold nanoparticles. The results showed the nonlinear optical properties of these samples.

Keywords : instrumentation; nonlinear refractive index; Z-scan; nanoparticles.

        · text in English     · English ( pdf )