RIVERO-RAMIREZ, D. C. et al. Automation of the acquisition and processing of data from thermoluminescence dosimeter readers. J. appl. res. technol [online]. 2005, vol.3, n.2, pp.118-124. ISSN 2448-6736.

The thermoluminescence dosimeter (TLD) readers used in the Laboratory of External Dosimetry of the Center of Protection and Hygiene from the Radiations are old-fashioned. The data obtained by this equipment must be manually written down in order to subsequently, process them and these data do not remain registered in any memory for future analysis. This paper describes the design and creation of a system that connects these TLD readers to a personal computer. The system includes a data acquisition card with several analogue input channels, counters, timers as well as digital inputs/outputs and software, in order to control the acquisition and processing of the data.

Keywords : Thermoluminescence; Dosimeter; Data Acquisition; Dosimetry; Dose Evaluation.

        · abstract in Spanish     · text in English

 