Development and validation of an analytical method for the determination of lead isotopic composition using ICPQMS
M. T. RodríguezSalazar1, 2, O. Morton Bermea3*, E. HernándezÁlvarez3, M. E. GarcíaArreola4, M. T. OrtuñoArzate5
1 Posgrado en Ciencias de la Tierra, Universidad Nacional Autónoma de México, Ciudad Universitaria, Del. Coyoacán, 04510, Mexico City, Mexico.
2 Facultad de Química, Universidad Nacional Autónoma de México, Ciudad Universitaria, Del. Coyoacán, 04510, Mexico City, Mexico.
3 Instituto de Geofísica, Universidad Nacional Autónoma de México, Ciudad Universitaria, Del. Coyoacán, 04510, Mexico City, Mexico. *Corresponding author: omorton@geofisica.unam.mx
4 Instituto de Geología, Universidad Autónoma de San Luis Potosí México, San Luis Potosí, Mexico.
5 Centro Nacional de Investigación y Capacitación Ambiental, Mexico City, Mexico.
Received: April 21, 2009
Accepted: February 17, 2010
Resumen
Este trabajo reporta un método para la determinación precisa y exacta la composición isotópica de Pb en matrices ambientales y geológicas mediante ICPQMS. Se analizaron réplicas de tres materiales de referencia (AGV2, SRM 2709 y JSO1), por medio de ICPQMS con el objeto de evaluar la calidad del procedimiento analítico. La discriminación de masas para Pb/Pb fue evaluado usando el material de referencia NIST 981, y esta corrección fue aplicada a los materiales de referencia antes mencionados, para obtener la exactitud del método analítico desarrollado. Se aplicó una estandarización interna con 205Tl para corregir la deriva instrumental. La composición isotópica obtenida para los materiales de referencia es comparable con los valores reportados (JSO1 and SRM 2709) y certificados (AGV2). Los valores de exactitud de los resultados obtenidos varían en promedio entre 0.002% para 207Pb/204Pb y 0.825% para 207Pb/206Pb. Estos resultados indican que el método ICPQMS es confiable en términos de precisión y exactitud para la determinación de relaciones isotópicas de Pb en matrices de suelos y geológicas.
Palabras clave: ICPQMS, composición isotópica de plomo, metodología analítica, material de referencia.
Abstract
This work reports a method for the precise and accurate determination of Pb isotope composition in soils and geological matrices by ICPQMS. Three reference materials (AGV2, SRM 2709 and JSO1) were repeatedly measured, using ICPQMS instruments in order to assess the quality of this analytical procedure. Mass discrimination was evaluated for Pb/Pb with Pb isotope reference material NIST SRM 981, and the correction applied to the above mentioned reference materials to achieve good accuracy of the analytical methodology. An internal standardization using 205Tl was employed to avoid instrumental drift. The measured isotope compositions of all analyzed reference material were comparable with the reported (JSO1 and SRM 2709) and certified values (AGV2). The accuracy results agreed in average within 0.002% for 207Pb/204Pb and 0.825% for 207Pb/206Pb. These data corroborate that ICPQMS results to be reliable in terms of accuracy and precision, for the determination of the Pb isotopic compositions in soils and geological matrices.
Key words: ICPQMS, lead isotope composition, analytical methodology, reference material.
Introduction
Lead has four naturallyoccurring isotopes: 204Pb, 206Pb, 207Pb, 208Pb. 204Pb is not radiogenic. The isotopic composition of Pb is commonly expressed as the ratios 206Pb/207Pb or 208Pb/206Pb. 206Pb/207Pb may be determined precisely analytically and the abundances of these isotopes are relatively important. The less abundant 204Pb is often used as a reference in geochronological studies because it is stable and its abundance is constant.
Another practical application of Pb isotope ratios is tracing sources of lead in the environment. The isotopic compositions characteristic of different types of Pbcontaining minerals aqre affected by the interaction of a number of processes, including radioactive decay of U and Th to Pb, the relative proportion of UThPb in the system, and mixing of Pb from different sources. Thus the isotopic relations are distinctive of different natural sources of Pb. When the contamination is due to a single process, the pollutant will have the isotopic signature of its source (Wong & Li 2004; Komárek et al. 2008; Cheng & Hu 2010).
Thermal ionization mass spectrometry (TIMS) is still regarded as the definitive methodology for lead isotope measurements. However, the relatively high cost of instrumentation and the time consuming chemical pretreatment of the sample prior to analysis limited the routine use of TIMS techniques in environmental studies where a large number of samples are analyzed. In the last ten years the utility of ICPQMS in Pb isotopic studies has been explored and the development of sensitive and highly precise analytical techniques based on ICPQMS has allowed the determination of isotopic relations of Pb in environmental and geologic materials with excellent results (Farmer, et al. 1996; Moor et al., 1996; Monna et al., 1997; Shotyk et al., 1998; Lima et al., 2005).
The main advantages of this technique include fast sample preparation, lower sample analysis cost as compared with TIMS, and instrument robustness (Townsend et al., 1998; Vanhaecke et al., 1998; AlAmmar and Barnes, 2001; Marguí et al., 2007). The accessibility of ICPQMS techniques has led to a higher number of Pb isotopic studies in recent years (Weiss et al., 2000; Mukal et al., 2001; Gallon et al., 2006; Marguí et al., 2007). However, this technique has been found to be of limited precision and accuracy owing to the design and operation of quadropole mass analyser (Townsend et al., 1998; Vanhaecke et al., 1998; Marguí et al., 2007).
The instrumental mass bias phenomenon affects the accuracy and precision of ICPQMS analysis (Vanhaecke et al., 1998; Al Ammar and Barnes, 2001). This phenomenon occurs when an instrument does not have the same sensitivity for different masses due to differences in the ion transmission. A suitable mass bias correction aims to improve accuracy and precision of the measurements. This correction involves:
1) Determination of dead time of the ion counting detection systems,
2) Mass bias evaluation using the Pb/Pb ratio of SRM 981 solution.
3) Application of the mass bias factor obtained for the reference material and/or experimental data for real samples.
Detector dead time is the time required for the detection and electronic handling of an ion pulse. The choice of the optimum dead time value is based on measuring the isotope ratio at different Pb concentrations. The data were reintegrated using different dead time values and the normalized ratio (experimental data/ certified data) was plotted versus the dead time for each of the different concentration solutions. The optimum dead time was determined to be the point at which the different concentration lines intercepted, i.e., the dead time for which the normalized isotope ratio has value more about to 1 and this parameter is independent of Pb concentration.
An internal standardization is employed to correct for instrumental drift. Thallium and bismuth are used as internal standards allowing longer acquisition periods for isotope ratio measurement of Pb as analyte (Quétel et al., 1997; Adgate et al., 1998; Barton et al., 2000; Ettler et al., 2004; Marguí et al., 2007). Lead isotopic composition of an environmental reference material for interlaboratory comparison tests using ICPQMS have been published by Aung et al. (2004). Tanimizu and Ishikawa (2006) reported Pb isotope ratios of rock reference samples from the Geological Survey of Japan using MCICPMS.
The purpose of this study is to provide a rapid and precise analytical method to determine lead Pb isotopic composition using ICPQMS in environmental and geologic matrices in a routine way. The method was validated via analysis of reference materials: AGV2 (USGS), SRM 2709 San Joaquin Soil (NIST) and JSO1 (Geological Survey of Japan). Furthermore, the results also add new data to the set of basic values of lead isotopic compositions, which will be helpful for further research in the field of environmental monitoring and assessment.
Experiment
The isotopic measurements were performed with two Model X, Series 2 ICPQMS spectrometers by Thermo Electron Corporation. One instrument was located at the Geological Institute of the Universidad Autónoma de San Luis Potosi, Mexico (UASLP) and the other at Centro Nacional de Investigacion y Capacitación Ambiental (CENICA), Mexico City, Mexico. The measurement conditions are summarized in Table 1.
Both instruments used Ni cones, Meinhard nebulizers and cooling camera. The instrument settings were optimized for maximum sensibility and minimum signal fluctuation with a multielemental tuning solution. The optimal dead time was selected for both instruments as 35 ns. Fig. 1 shows the dead time determination for 208Pb/206Pb.
The optimal instrumental dead time was adjusted for all analyzed isotope ratios using Pb solutions of different concentrations (5, 10, 20, 50 and 100 ppb), prepared from a Pb stock solution of NIST SRM 981 by dissolving about 1 g of Pb metal wire in 20 ml HNO3 15% transferred into a 1 1 flask and brought to volume with MilliQwater (Townsend et al., 1998). In order to determine isotope ratios accurately, the mass bias factor was evaluated using the Pb/Pb ratio of the SRM 981 solution.
Standard solutions and reference material for mass spectrometry were prepared with purified water obtained from a 18 M'Ωgrade Millipore system. Ten replicate measurements were performed for all isotopic masses (205Tl 204Pb 206Pb 207Pb 208Pb)
The experimental procedure was evaluated for three standard reference materials: AGV2 (USGS), SRM 2709 San Joaquin Soil (NIST) and JSO1 (Geological Survey of Japan). The same digestion procedure was applied to all reference materials. About 500 mg of reference material were digested using a mixture of acids (10 ml HNO3 concentrated + 3 ml HF concentrated +5 ml H2O2 (30%) in a CEM microwave, MARS XPress digestion system. After the digestion, a minimal residue was obtained from the AGV2 reference material compared to the digested soils reference material. After cooling, sample digests were filtered through a Whatman 42 filter, transferred into a 100 ml flask and brought to volume with MilliQwater.
The raw mass data were processed for subtraction of blank and corrected for the internal standard ratio signal (205Tl) and for mass bias. The impact of 204Hg on the measurement of 204Pb was evaluated and found to be essentially negligible, because of the low total Hg concentration in the analyzed samples.
Results and Discussion
Lead isotope ratios obtained for all analyzed reference materials are summarized in Table 2. The precision of the isotopic ratio measurement, was calculated in terms of relative standard deviation (%RSD) of two replicates. The RSD was better than 1.3% for all obtained values (206Pb/204Pb in AGV2 at CENICA). Reference material NIST 2709 measured at UASLP shows the best precision (0.06 % RSD for 208Pb/206Pb). Accuracy was assessed in terms of % difference between reported and experimental results. The results obtained with both ICPQMS instruments show good agreement with the reported values. Results at UASLP differed from the standard values by 0.002% to 0.625% at UASLP and by 0.009% to 0.825% at CENICA. The peak accuracy in terms of 207Pb/204Pb ratios was obtained for the AGV2 standard at UASLP (0.002%), and in terms of 208Pb/206Pb ratios for the NIST 2709 standard at CENICA (0.009%). There is no indication of a trend that might suggest a higher accuracy for any specific isotopic ratio. There is no observable dependence of the quality of the results on the analyzed matrix.
Conclusions
The experimental results reported in this paper suggest that good precision and accuracy may be achieved by using ICPQMS to determine lead isotope ratios provided that the experimental procedure is carefully designed and appropriate corrections for blank, instrumental drift, dead time, interference, and mass bias are used. The technique may be used in routine analysis of soils and geological samples for environmental monitoring and evaluation.
Precision values ranged between 0.06% and 1.3% RDS, and accuracy varied between errors of 0.002% to 0.825%. No dependence on isotopic ratios or on the type of matrix was detected. The methodology is reliable and well suited to the kind of facilities available in most Mexican laboratories. ICPQMS is a competitive, promising technique of lead isotope ratio analysis, because of its low cost, easy sample preparation and widely availably compared with other mass spectrometric techniques.
Acknowledgements
This study was performed with financial support of DGAPA (Dirección General de Personal Académico, UNAM), project IN117405. The authors thank Mercedes Reyes (CENICA) and the LAFQAUNAM staff (Margarita Gutiérrez, Gerardo Martínez, Agueda Ceniceros, Hedgar Hernández) for providing analytical facilities. RodríguezSalazar gratefully acknowledges a grant from PASPA (Programa de Apoyo a la Superación del Personal Académico, DGAPA, UNAM). We thank Laura Beramendi for a critical revision of the manuscript.
Bibliography
Adgate, J. L., G. G. Rhoads and P. J. Lioy, 1998. The use of isotope ratios to apportion sources of lead in Jersey City, NJ, house dust wipe samples. Sci. Total Environ. 221, 171180. [ Links ]
AlAmmar, A. S. and R. M. Barnes, 2001. Improving isotope ratio precision in inductively coupled plasma quadrupole mass spectrometry by common analyte internal standardization. J. Anal. At. Spectrom. 16, 327332. [ Links ]
Aung, N. N., T. Uryu and J. Yoshinaga, 2004. Lead Isotopic Compositions of Environmental Certified Reference Materials for an InterLaboratory Comparison of Lead Isotope Analysis. Anal. Sci. 20, 195198. [ Links ]
Barton, H., Z. Zachwieja, S. D'Ilio and S. Caroli, 2000. Application of routine estimation of Pb isotopic ratios by inductively coupled plasma mass spectrometry for studying the Pb origin in hair of children living in polluted areas. A pilot study. Microchem. J. 67, 21-30. [ Links ]
Cheng, H. and Y. Hu. Lead (Pb) isotopic fingerprinting and its applications in lead pollution studies in China: A review. Environ. Pollut. 2010, doi:10.1016/ j.envpol.2009.12.028 [ Links ]
Ettler, V., M. Mihaljevic and M. Komárek, 2004. ICPMS measurements of lead isotopic ratios in soils heavily contaminated by lead smelting: tracing the sources of pollution. Anal. Bioanal. Chem. 378, 311317. 116 Geofis. Int. 49 (3), 2010 [ Links ]
Farmer, J. G., L. J. E ades, A. B. Mackenzie, A. Kirika and T. E. BaileyWatts, 1996. Stable Lead Isotope Record of Lead Pollution in Loch Lomond Sediments since 1630 A.D. Environ. Sci. Technol. 30, 30803083. [ Links ]
Gallon, C., A. Tessier, C. Gobeil and R. Carignan, 2006. Historical Perspective of Industrial Lead Emissions to the Atmosphere from a Canadian Smelter. Environ. Sci. Technol. 40, 741747. [ Links ]
Komárek, M., V. Ettler, V. Chrastny and M. Mihaljevic, 2008. Lead isotopes in environmental sciences: A review. Environ. Int. 34, 562577. [ Links ]
Lima, A.L., B. A. Bergquist, E.A. Boyle, M. K. Reuer, F. O. Duda, C. M. Reddy and T.I. Eglinton, 2005. Highresolution historical records from Pettaquamscutt River basin sediments: 2. Pb isotopes reveal a potential new stratigraphic marker. Geochim. Cosmochim. Acta 69, 18131824. [ Links ]
Marguí, E., M. Iglesias, I. Queralt and M. Hidalgo, 2007. Precise and accurate determination of lead isotope ratios in mining wastes by ICPQMS as a tool to identify their source. Talanta 73, 700709. [ Links ]
Monna, F., J. Lancelot, I. W. Croudace, A. B. Cundy and J. T. Lewis, 1997. Pb Isotopic Composition of Airborne Particulate Material from France and the Southern United Kingdom: Implications for Pb Pollution Sources in Urban Areas. Environ. Sci. Technol. 31, 22772286. [ Links ]
Moor, H. C., T. Schaller and M. Sturm, 1996. Recent Changes in Stable Lead Isotope Ratios in Sediments of Lake Zug, Switzerland. Environ. Sci. Technol. 30, 29282933. [ Links ]
Mukai, H., A. Tanaka, T. Fujii, Y. Zeng, Y. Hong, J. Tang, S. Guo, H. Xue, Z. Sun, J. Zhou, D. Xue, J. Zhao, G. Zhai, J. Gu and P. Zhai, 2001. Regional Characteristics of Sulfur and Lead Isotope Ratios in the Atmosphere at Several Chinese Urban Sites. Environ Sci Technol. 35, 10641071. [ Links ]
Quétel, CR, B. Thomas, O. F. X. Donard and F. E. Grousset, 1997. Factorial optimization of data acquisition factors for lead isotope ratio determination by inductively coupled plasma mass spectrometry. Spectrochim. Acta Part B 52, 177187. [ Links ]
Shotyk, W., D. Weiss, P. G. Appleby, A. K. Cheburkin, R. Frei, M. Gloor, J. D. Kramers, S. Reese and W. O. Van Der Knaap, 1998. History of Atmospheric Lead Deposition Since 12,370 14C yr BP from a Peat Bog, Jura Mountains. Switzerland Science 281, 16351640 [ Links ]
Tanimizu, M. and T. Ishikawa, 2006. Development of rapid and precise Pb isotope analytical techniques using MCICPMS and new results for GSJ rock reference samples. Geochem. J. 40, 121123. [ Links ]
Townsend, A., Z. Yu, P. Mcgoldrick and J. A. Hutton, 1998. Precise lead isotope ratios in Australian galena samples by high resolution inductively coupled plasma mass spectrometry. J. Anal. At. Spectrom. 13, 809813. [ Links ]
Vanhaecke, F., G. Wannemacker, L. Moens, R. Dams, C. Latkoczy, T. Prohaska and G. Stingeder, 1998. Dependence of detector dead time on analyte mass number in inductively coupled plasma mass spectrometry. J. Anal. At. Spectrom. 13, 567571. [ Links ]
Weiss, D., E. A. Boyle, V. Chavagnac, M. Herwegh and J. Wu, 2000. Determination of lead isotope ratios in seawater by quadrupole inductively coupled plasma mass spectrometry after Mg(OH)2 coprecipitation. Spectrochim. Acta Part B 55, 363374. [ Links ]
Wong, C. S. C., and X. D. Li, 2004. Pb contamination and isotopic composition or urban soils in Hong Kong. Sci. Total Environ. 319, 185195. [ Links ]